269 020 584 E 0806
Additives & Instruments
Measure what you see.
BYK-Gardner GmbH · Lausitzer Strasse 8 · 82538 Geretsried · Germany · Tel. +49-8171-3493-0 · Fax +49-8171-3493-140 · www.byk.com/instruments
Please note detailed operating instructions in English on CD-ROM.
ENGLISH
Operation
mode wheel:
• Switch on press
• Show menu press
• Select function scroll
• Activate function press
esc button:
• Change between menu
and measurement mode
std button:
• Short-cut to measure standard
Operate button:
• Switch on, select and measure
How to Calibrate?
Black/white calibration
• Select Setup – Calibrate – Black/White Calibration
• Place unit on black trap and confirm OK
• Place unit on white standard and confirm OK
Black/white calibration is requested every 3 months.
Cyan check
After a black and white calibration the instrument must be checked on the
Cyan standard.
• Place unit on Cyan standard and confirm OK
The Cyan check is requested every 30 days. Make sure to perform it in
laboratory environment (Temperature 20 - 25°C, Humidity 30 - 60%).
operate
How to measure Differences?
• Select a stored standard, use
„AUTO standard“ or measure a standard
• Input standard name:
Change character ==> turn scroll wheel
Move to next digit ==> press scroll wheel
Move one digit back ==> press esc
• Input number of readings
• Place instrument on standard
• Press operate to take a reading
• Copy predefined settings for data display
• Continue with sample readings
Note: All color data inclusive spectral data are
saved and transferred to software.
• First page displays color
differences
• Scroll to second page for sparkle
and graininess results*
* Only available with BYK-mac i