10.6.2Auto standard
When the “Auto Std” function is activated, the
standard which corresponds most closely to the
sample is automatically chosen from the set of
stored standards, and the differences are displayed.
This function can be toggled on and off. It is only
active in the Difference and Pass/Fail modes.
From the “Options” menu, use the
button to move the cursor to the “Setup” option. In
the submenu, place the cursor on “Auto Std” and
press the Operate button. The toggled status is
shown on the display, e.g.:
Auto Std on!
(or off). After a measurement has been performed in
the Difference mode, the message “Auto Std”
appears at lower left, in the next-to-last line. The
name of the automatically selected standard
appears at lower right, in the last line.
The “Std” button is not active in this mode.