Exact calibration is only possible
with perfect standards. If the
condition on the standards should
appear doubtful to you on the basis
of a visual inspection or a
measurement error, we will be glad
to check the standards.
7.3 Changing Calibration Values
The values of the supplied
calibration standards are stored in
the measuring unit. When
automatically calibrating, this data
is allocated to the respective
standard in the holder.
In some cases, it may become
necessary to enter the data for a
new calibration standard. This
might be the case if for example
the standard used so far is
damaged or scratched.
Only use original standards of
the manufacturer to ensure exact
calibration.
The function CHANGE
CALIBRATION VALUES serves to
enter new calibration values.
To calibrate the unit, push it into the
holder until the stop. Now switch
on the unit using the mode button.
By actuating the mode button
several times, skip calibration until
CHANGE CAL. VALUES appears
(see fig. 12).
7 Calibration
Fig. 12: Changing calibration values
CHANGE CAL.CHANGE CAL.
CHANGE CAL.CHANGE CAL.
CHANGE CAL.
VALUES VALUES
VALUES VALUES
VALUES
--
--
-
>>
>>
>
operateoperate
operateoperate
operate
CHANGE CAL.VALUECHANGE CAL.VALUE
CHANGE CAL.VALUECHANGE CAL.VALUE
CHANGE CAL.VALUE
20°20°
20°20°
20°
--
--
-
>>
>>
>
operateoperate
operateoperate
operate
20°VALUE: 0091.720°VALUE: 0091.7
20°VALUE: 0091.720°VALUE: 0091.7
20°VALUE: 0091.7
--
--
-
>>
>>
>
operateoperate
operateoperate
operate
^^
^^
^
NEW VALUE: 0092.3NEW VALUE: 0092.3
NEW VALUE: 0092.3NEW VALUE: 0092.3
NEW VALUE: 0092.3
CONFIRM CONFIRM
CONFIRM CONFIRM
CONFIRM
--
--
-
>>
>>
>
operateoperate
operateoperate
operate
CHANGE CAL.VALUECHANGE CAL.VALUE
CHANGE CAL.VALUECHANGE CAL.VALUE
CHANGE CAL.VALUE
HlinHlin
HlinHlin
Hlin
--
--
-
>>
>>
>
operateoperate
operateoperate
operate
operate 5x
mode
CHANGE CAL.VALUECHANGE CAL.VALUE
CHANGE CAL.VALUECHANGE CAL.VALUE
CHANGE CAL.VALUE
YY
YY
Y
--
--
-
>>
>>
>
operateoperate
operateoperate
operate
change marked figure:
operate
confirm marked figure:
mode
mode operate
operate:
take over new value
mode:
retain old value
operate: continue as
with 20° value
operate: continue as
with 20° value