42
Geometry 20°
Measured area 9 x 9 mm (0.4 x 0.4")
Light source Tungsten filament lamp (2.5 V/60 mA)
Detector Silicon photoelement
Spectral evaluation approx. CIE luminosity function y (2°) under
CIE-illuminant C
Memory capacity 999 measurements, backed by a lithium battery,
maintains data for approximately 5 years
Interface serial (RS232C)
Power supply 2 alkaline batteries 1,5V (Mignon AA), approx.
5000 measurements or 2 NC or NH batteries
1,2V (Mignon AA), approx. 1500 measurements
(Service life can differ, depending on brand.)
Dimensions (WxLxH)
measuring unit 43 x 141 x 72 mm (1.7 x 5.5 x 2.8")
protective holder 55 x 148 x 65 mm (2.2 x 5.8 x 2.6")
Weight:
measuring unit 480 g (1lb)
holder 120 g (0.3 lb)
Temperature range – 10 °C … + 60 °C storage (15 … 160 F)
+ 15 °C … + 40 °C operation (60 … 100 F)
Gloss
Measuring range 0 –180 gloss units (GU)
Repeatability 0,5 GU
Reproducibility 1 GU in a range between 0 and 100 GU
Haze
Measuring range 10 … 500 units (Hlog)
Repeatability* ± 1 unit (Hlog)
Reproducibility* ± 7 units (Hlog)
* Due to logarithmic scaling the indications relate to
values greater than 10 units
Specifications subject to change without notice.