Errors and warning messages 22
Error Thickness Operating error: improper application, raising
before measurement is complete, or calibration
error. Repeat the procedure.
If the error is shown repeatedly despite
correct operation and calibration, please contact
our customer service department.
Please observe the instructions on cleaning standards in the section on
Calibration.
Fluctuations in measurement values
No. Check how high the
deviations are on the sample itself.
Calibration not correct:
Recalibrate and clean the
standard if necessary.
No: In this case, major deviations
are possible.
Please contact our Customer Ser-
vice department.
Do not attempt to make any repairs yourself! If a malfunction occurs on
your measuring device, our Customer Service department will be happy
to help you as quickly as possible.
Was the same point on the
sample used for all
measurements?
Yes. It may help to test the
calibration with an additional
standard if one is available.
Calibration correct: Is the test
surface completely even and
does the measuring device have
good contact with the sample?
Yes: Device defective